MANUFACTURED JULY, 2023
Cell-IQ™ Series 5.8 cu.ft. CO2 Laboratory Incubator MCO-170AICUVHL-PA
Cell Culture Incubators Capabilities
As a crucial piece of laboratory equipment, PHCbi brand cell culture incubators are designed to ensure highly accurate, reproducible chamber conditions. Our cell culture incubators provide a contamination-free environment while allowing precise CO2 or CO2/O2 control – essential for in vivo environment replication.
Time Saving Decontamination
An integral decontamination system using hydrogen peroxide (H2O2) vapor safely cleans the interior chamber in less than three hours after initiation. This system is safe, efficient and effective against contaminants that typically enter the incubator during door openings. A 100% kill rate with at least six log reduction is documented against major pathogens. Following cleaning, the SafeCell™ UV light reduces H2O2 to harmless water vapor and the incubator can be immediately placed back into service.
Uniform and Consistent Environment
The patented Direct Heat and Air Jacket system, IR2 sensor, SafeCell™ UV and inCu-saFe® work together to ensure precise temperature, CO2 levels and H2O2 contamination control respectively for a precise and repeatable environment.
Rapid CO2 Recovery with IR2 Sensor
Proprietary IR2 sensor provides superior CO2 control with fast recovery without overshoot even following multiple door openings.
Contamination Control with SafeCell UV and InCu-saFe
PHCbi’s inCu-saFe combines the germicidal properties of elemental copper with the corrosion and discoloration resistance of stainless steel to deliver passive contamination control throughout the chamber interior. The SafeCell UV lamp provides safe, active and passive contamination control of circulated air and the humidity water reservoir. SafeCell scrubs and decontaminates circulated chamber air without affecting cell cultures in situ.
Color LCD Display
The high contrast color LCD display allows for easy, touchscreen programming of temperature, CO2 and other internal conditions of your incubator even with gloved hands.
USB Log Function for Data Transfer and Storage
A standard USB port simplifies the storage or transfer of logged data.
EXTERIOR DIMENSIONS | 24.4″ W x 28.7″ D x 35.6″ H (620 x 730 x 905 mm) |
---|---|
INTERIOR DIMENSIONS | 19.3″ W x 20.6″ D x 26.2″ H (490 x 523 x 665 mm) |
INTERIOR VOLUME | 5.8 cu.ft. (165 L) |
DATA ACQUISITION | Optional analog on cloud-based monitoring |
OUTER DOOR | Painted steel |
SHELVES | 18.5″ x 17.7″ (470 x 450 mm) |
ACCESS PORT | 1: Back side (1.2″ (30 mm) diameter) |
INSULATION | Extruded polystyrene foam |
AIR FILTER | 0.3 um, Efficiency: 99.97 % or higher |
UV LAMP | 4W x 1 (ozone-free emission) |
NET WEIGHT | 176 lbs (80 kg) |
TEMPERATURE CONTROL RANGE | Ambient temperature +º5C to +50ºC |
TEMPERATURE UNIFORMITY | ±0.25ºC (ambient temperature: 23ºC, setting: 37ºC, CO2: 5 %, no load) |
MAXIMUM POWER CONSUMPTION | Max. 376 W |
CO2 CONTROL RANGE | 0% to 20% |
CO2 CONTROLLER | PID Control System |
CO2 GAS CONNECTION | 0.157 inch to 0.36 inch (4 mm to 6 mm) diameter tube can be connected |
CO2 GAS PRESSURE | 0.03 MPa(G) (0.3kgf/cm2(G),4.3 psi(G)) |
CO2 VARIATION | ±0.15% (ambient temperature: 23ºC, setting: 37ºC, CO2: 5%, no load) |
CO2 SENSOR TYPE | Dual filter infrared sensor |
DATA OUTPUT | USB |
DECONTAMINATION METHODS | SafeCell ultraviolet light. Hydrogen peroxide (H2O2) vapor system |
CONTAMINATION CONTROLS | InCu-SaFe copper enriched stainless steel interior and shelves, SafeCell UV light |
DECONTAMINATION CYCLE TIME | < 3.0 Hours |
DISPLAY | LCD touchscreen |
HEATING SYSTEM | Direct Heat and Air Jacket System |
HUMIDIFYING SYSTEM | Natural evaporation with humidifying pan, 95 %R.H.±5 %R.H. |
INTERIOR SURFACES | InCu-saFe stainless steel with copper enriched alloy |
Reviews